Command reference
R&S
®
ZNA
1468User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> *RST: OFF (0)
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Coupled Settings" on page 696
[SENSe<Ch>:]PULSe:DUTYcycle <Percentage>
Defines the duty cycle of the generated pulses.
This setting only takes effect, if:
●
Internal Pulse Control is enabled ([SENSe<Ch>:]PULSe:INTernal[:STATe]
ON)
●
Fixed Duty Cycle is enabled ([SENSe<Ch>:]PULSe:FXDCycle[:STATe] ON)
Note that currently the exact value of the duty cycle can not be guaranteed. However,
you can be sure that the specified value is not exceeded.
Suffix:
<Ch>
.
Channel number
Parameters:
<Pulse Duty Cycle> A value between 0 and 100.
Default unit: percent
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Fixed Duty Cycle/Duty Cycle" on page 696
[SENSe<Ch>:]PULSe:FXDCycle[:STATe] <Boolean>
Defines whether the analyzer firmware shall enforce a fixed duty cycle for all pulse
generators.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Fixed duty cycle for all pulse generators in channel <Ch>.
The pulses are defined using [SENSe<Ch>:]PULSe:PERiod
and [SENSe<Ch>:]PULSe:DUTYcycle.
OFF (0)
The generator-specific pulses are defined using [SENSe<Ch>:
]PULSe:PORT<Pt>:WIDTh and [SENSe<Ch>:]PULSe:
PORT<Pt>:DELay.
*RST: OFF
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
SCPI command reference