Command reference
R&S
®
ZNA
1474User Manual 1178.6462.02 ─ 29
Manual operation: See "PuMo In" on page 699
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:OUTPut:INVerted[:STATe] <Boolean>
If pulse generator <Pt> is output to PuMo In/Out <Pt> (see [SENSe<Ch>:]PULSe:
PORT<Pt>:EXTernal:OUTPut[:STATe]), this defines the polarity of the outgoing
pulses.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Boolean> *RST: OFF (0)
Options: R&S ZNA-B91
Manual operation: See "PuMo Out" on page 698
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:OUTPut[:STATe]
<Boolean>
If Internal Pulse Control is used, this enables/disables output of pulse generator <Pt>
at PuMo In/Out <Pt>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Boolean> *RST: OFF (0)
Options: R&S ZNA-B91
Manual operation: See "PuMo Out" on page 698
[SENSe<Ch>:]PULSe:PORT<Pt>[:STATe] <Boolean>
Activates or deactivates the internal pulse modulator at VNA port <Pt>.
Suffix:
<Ch>
.
Channel number
<Pt> VNA port number
Parameters:
<Boolean>
Options: R&S ZNA<Freq>B4<Pt>
Manual operation: See "Port" on page 698