1049User Manual 1178.6462.02 ─ 29
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> 1 (ON, true) if the test coupon is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Manual operation: See "Active" on page 732
CALCulate:FMODel:SFD:COUPon:TYPE <CouponType>
Sets/gets the coupon type to be measured for a fixture modelling with the SFD tool.
Parameters:
<CouponType> SYMMetric2x | OPEN1x | SHORt1x
Manual operation: See "Coupon Type" on page 731
CALCulate:FMODel:SFD:DIFFcfg
<SFDPortConfig>
Tells the SFD tool about the port ordering of the 2x Thru test coupon.
Parameters:
<SFDPortConfig> ODD | NON
ODD: odd ports are on the left and even ports are on the right
NON: ports 1 to N are on the left and ports N+1 to 2·N are on
the right
Manual operation: See "2x Thru Port Ordering" on page 740
CALCulate:FMODel:SFD:DUT:MEASure
For a fixture modeling with the ISD tool, this command allows to measure the DUT +
Fixture at the active ports (see CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe]).
Usage:
Event
Manual operation: See "Measure" on page 733
CALCulate:FMODel:SFD:DUT:MEASure:FILename <String>
For a fixture modeling with the EZD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
Manual operation: See "Load File" on page 733