Command reference
R&S
®
ZNA
1411User Manual 1178.6462.02 ─ 29
Manual operation: See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 708
[SENSe<Ch>:]CORRection:LOSS:AUTO <Activate>
Applies the Auto Length and Loss) function to the active trace of channel <Ch>.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Activate> ONCE
Required constant
Example:
*RST; :CORR:LOSS:AUTO ONCE
Reset the instrument and apply the "Auto Length and Loss"
function to the default trace (Trc1 in channel 1).
Usage: Setting only
Manual operation: See "Auto Length and Loss" on page 723
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:FREQuency <RefFreq>
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:FREQuency <RefFreq>
These commands define the reference frequencies f
1
and f
2
for the offset loss approxi-
mation.
The corresponding losses can be defined using [SENSe<Ch>:]CORRection:
LOSS<PhyPt>:OFFSet.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<RefFreq> Reference frequency
Range: Frequency range of the analyzer model.
Increment: 1 MHz
*RST: 1000000000 Hz (= 1 GHz)
Default unit: Hz
Manual operation: See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 708
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet <OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
SCPI command reference