Command reference
R&S
®
ZNA
1412User Manual 1178.6462.02 ─ 29
Parameters:
<OffsetLoss> Frequency-independent part of the offset loss
Range: -200 dB to +200 dB
Increment: 0.001 dB
*RST: 0 dB
Default unit: dB
Example: See [SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation: See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 708
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:STATe <Boolean>
Defines whether the second reference frequency f
2
is used for the offset loss approxi-
mation.
The second reference frequency and its associated loss value can be defined using
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:FREQuency and
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> OFF (0)
f
2
not used (default)
ON (1)
f
2
used
Manual operation: See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 708
[SENSe<Ch>:]CORRection:NFIGure[:STATe] <Boolean>
Disables or enables the Noise Figure Calibration for the related channel.
Suffix:
<Ch>
.
Parameters:
<Boolean> OFF - Disable calibration
ON - (re-)enable calibration
*RST: OFF (calibration disabled, no noise figure calibra-
tion available after a reset)
Options: R&S ZNA-K30
SCPI command reference