CALCulate:FMODel:EZD:IMPedance:REFerence <Value>
Defines the reference impedance for the impedance correction of the EZD tool (see
CALCulate:FMODel:EZD:IMPedance).
Parameters:
<Value> Default unit: Ohm
Options: R&S ZNA-K210
Manual operation: See "Impedance Correction Settings" on page 742
CALCulate:FMODel:EZD:PORT:ORDer <PortOrder>
Tells the EZD tool about the port ordering of the test fixture.
Parameters:
<PortOrder> ODD | NON
ODD: odd ports are on the left and even ports are on the right
NON: ports 1 to N are on the left and ports N+1 to 2·N are on
the right
Options: R&S ZNA-K210
Manual operation: See "Port Order" on page 741
CALCulate:FMODel:EZD:PRESet
Restores the global default settings of the EZD tool.
Usage:
Event
Options: R&S ZNA-K210
Manual operation: See "Reset to Default" on page 730
CALCulate:FMODel:EZD:RUN:RUN
Runs the EZD tool.
Before executing this command, make sure that:
●
The test coupon measurement (using CALCulate:FMODel:EZD:COUPon:
MEASure) finished successfully, or the test coupon data were successfully loaded
from file (using CALCulate:FMODel:EZD:COUPon:MEASure:FILename
●
The measurement of DUT + test fixture (using CALCulate:FMODel:EZD:
COUPon:MEASure) finished successfully.
The resulting Touchstone files are written to C:\Users\Public\Documents\Rohde-
Schwarz\ZNA\Embedding. If result files with the same name already exist, they are
overwritten.
Usage:
Event
Options: R&S ZNA-K210