Command reference
R&S
®
ZNA
1461User Manual 1178.6462.02 ─ 29
IN
The ports are for IF input (but not for IF output).
Note that direct generator/receiver access cannot be used in this
case.
OUT
The ports are for IF output (but not for IF input).
Usage: Setting only
Options: R&S ZNA-B26
Manual operation: See "Rear Out" on page 678
[SENSe<Ch>:]PATH<Pt>:MEASurement:DIRectaccess?
[SENSe<Ch>:]PATH<Pt>:REFerence:DIRectaccess?
Queries the direct generator/receiver access path.
To define the direct generator/receiver access path, use [SENSe<Ch>:]PATH<Pt>:
DIRectaccess.
Note that direct generator/receiver access and direct IF input cannot be used at the
same time. Direct generator/receiver access is only used if [SENSe<Ch>:
]PATH<Pt>:IFSWitch is set to OFF or OUT.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Usage: Query only
Options: R&S ZNAxx-B16
Manual operation: See "Receiver Input" on page 678
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFINpath?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFINpath?
Query for the input paths of Direct IF Access Input / Output connectors IF Reference
<Pt> and IF Meas <Pt> in channel <Ch>.
To set the input paths, use [SENSe<Ch>:]PATH<Pt>:IFINpath.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Usage: Query only
Options: R&S ZNA-B26
Manual operation: See "Rear Input Path" on page 679
SCPI command reference