Command reference
R&S
®
ZNA
1118User Manual 1178.6462.02 ─ 29
<Seg> Number of the ripple limit range.
Parameters:
<Boolean> ON | OFF - Limit check on or off. A result is available even if the
limit check is disabled; see example for CALCulate<Chn>:
RIPPle:SEGMent<Seg>:RESult?.
*RST: n/a (no ripple limit line defined after a reset)
Example: See CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:
STARt
Manual operation: See "Range List" on page 490
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STARt
<FreqPowTime>
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STOP <StimValue>
These commands change the start or stop stimulus values (i.e. the smallest or largest
stimulus values) of a ripple limit range. A range must be created first to enable these
commands (e.g CALCulate<Chn>:RIPPle:DATA). See Chapter 4.4.1.2, "Rules for
ripple test definition", on page 171.
Tip: To define several ripple limit ranges with a single command, use
CALCulate<Chn>:RIPPle:DATA.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Number of the ripple limit range.
Parameters:
<StimValue> Stimulus values (unitless) confining the ripple limit range.
The unit is adjusted to the sweep type of the active channel
([SENSe<Ch>:]SWEep:TYPE).
Range: Virtually no restriction for ripple limit ranges.
*RST: n/a (no ripple limit line defined after a reset)
Default unit: NN
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:SEGM:STIM:STAR 1GHZ; STOP 2.5 GHZ; :
CALC:RIPP:SEGM:LIM 5
Change the range to a stimulus range between 1 GHz and 2.5
GHz and a limit of 5 dB.
CALC:RIPP:SEGM:STAT OFF
Disable the limit check in the modified stimulus range.
Manual operation: See "Range List" on page 490
CALCulate<Chn>:RIPPle:SOUNd[:STATe] <Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded ripple limit.