Command reference
R&S
®
ZNA
1576User Manual 1178.6462.02 ─ 29
Parameters:
<RecCalPower> Reference receiver cal power
Range: Depending on source power range of the analyzer
and the power range of the external power meter.
*RST: 0 dBm
Default unit: dBm
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire].
Manual operation: See "Reference Receiver Cal Power" on page 631
SOURce:POWer:CORRection:PPOWer:PATTenuation
<SrcAtt>
Defines the source step attenuation the R&S ZNA uses during the first calibration
sweep of the source power calibration ("Reference Receiver Cal Power"). This attenu-
ation value is ignored if the R&S ZNA is set to use the calculated port power result
(SOURce:POWer:CORRection:PSELect CPOW).
Parameters:
<SrcAtt> Attennuation value
Range: 0 to 70 dB
Increment: 10 dB
*RST: 0 dB
Default unit: dB
Manual operation: See "Reference Receiver Cal Power" on page 631
SOURce:POWer:CORRection:PSELect <CalPower>
Defines how to define the source power the R&S ZNA uses to perform the first calibra-
tion sweep of the source power calibration ("Reference Receiver Cal Power").
Parameters:
<CalPower> CPOWer | PPOWer
CPOWer
Calculated port power result
PPOWer
Predefined port power value, see SOURce:POWer:
CORRection:PPOWer
*RST: CPOWer
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire].
Manual operation: See "Reference Receiver Cal Power" on page 631
SCPI command reference