EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1336

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1336User Manual 1178.6462.02 ─ 29
<LocalIndex> The port index, if required.
For NONe and RLO the index is ignored.
Options: R&S ZNA-K8
Manual operation: See "Local" on page 903
[SENSe:]CONVerter<Port>:RFLoss <Loss>
Defines the cable loss at the source port or converter <Port>.
Suffix:
<Port>
.
Converter port number (= physical VNA port number)
Parameters:
<Loss> Loss value
Default unit: dB
Options: R&S ZNA-K8
Manual operation: See "Cable & Splitter Loss" on page 907
[SENSe:]CONVerter<Port>:RFPort?
Returns the VNA source port that drives converter <Port>.
Always returns <Port>.
Suffix:
<Port>
.
Converter number
Usage: Query only
Options: R&S ZNA-K8
Manual operation: See "Source" on page 903
[SENSe:]CONVerter<Port>:RFSLope <Slope>
Suffix:
<Port>
.
Parameters:
<Slope> Default unit: DB/GHZ
Manual operation: See "Slope" on page 676
[SENSe:]CONVerter:SPLitter<Port>:LOPort <Local>[, <LocalIndex>]
Defines the source of the LO signal distributed via splitter <Port>.
"Splitter 2" is only useful, if 2 independent 2-port converter measurements shall be per-
formed.
SCPI command reference

Table of Contents

Related product manuals