Example: Suppose we have configured a setup with two converters ZVA-
Z110.
SOURce:POWer:CORRection:SLEVeling:FREQuency:STARt 75 GHz
SOURce:POWer:CORRection:SLEVeling:FREQuency:STOP 110 GHz
SOURce:POWer:CORRection:SLEVeling:FREQuency:POINts 101
SOURce:POWer:CORRection:SLEVeling:POWer:STARt -25
SOURce:POWer:CORRection:SLEVeling:POWer:STOP 0
SOURce:POWer:CORRection:SLEVeling:POWer:STEP 0.5 DB
SOURce:POWer:CORRection:SLEVeling:POWer:ALC ON
SOURce:POWer1:CORRection:SLEVeling:ACQuire
Usage: Event
Options: R&S ZNA-K8
Manual operation: See "Start Cal Sweep" on page 604
SOURce:POWer<PhyPt>:CORRection:SLEVeling:DATA:CLEar
Deletes the leveling data for physical port <PhyPt>, if available.
Use SOURce:POWer:CORRection:SLEVeling:DATA:CLEar:ALL to delete the lev-
eling data for all ports.
Suffix:
<PhyPt>
.
Physical port number
Usage: Event
Options: R&S ZNA-K8
SOURce:POWer:CORRection:SLEVeling:DATA:CLEar:ALL
Deletes the leveling data for all physical ports.
Use SOURce:POWer<PhyPt>:CORRection:SLEVeling:DATA:CLEar to delete the
leveling data for a single port.
Usage:
Event
Options: R&S ZNA-K8
SOURce:POWer<PhyPt>:CORRection:SLEVeling:STATe?
Queries whether a leveling data set is available for physical port <PhyPt>.
Suffix:
<Ch>
.
Channel number
This suffix is ignored. Leveling data are valid for all channels in
the active setup.
Return values:
<Boolean>
Usage: Query only