Command reference
R&S
®
ZNA
1555User Manual 1178.6462.02 ─ 29
Options: R&S ZNA-K8
Manual operation: See "Available" on page 908
SOURce:POWer<PhyPt>:CORRection:SLEVeling:DELete
Removes the (global) leveling dataseet of port <PhyPt>.
Suffix:
<PhyPt>
.
Physical port number
Usage: Event
Options: R&S ZNA-K8
Manual operation: See "Remove" on page 908
SOURce:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:POINts
<NrOfPts>
Defines the number of frequency sweep points of the leveling grid for physical port
<PhyPt>.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<NrOfPts> Number of points
By default the R&S ZNA uses 201 sweep points. The higher the
number of "Sweep Points", the more accurate the leveling, but
also the longer it takes to create the leveling table). The number
of sweep points is limited to 100001.
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "Number of Points" on page 603
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STARt
<Leveling Tool Start Frequency>
SOURce:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STOP <FreqValue>
Defines the start and stop frequency of the leveling grid for physical port <PhyPt>.
By default the vector network analyzer uses the standard frequency range of the
selected "Meas Port".