After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<FreqValue> Start/stop value
Default unit: Hz
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "Start Frequency/Stop Frequeny" on page 603
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:ALC
<Compatible>
If checked, the power step size of the leveling grid for physical port <PhyPt> is fixed to
1 dB. The step size defined using SOURce:POWer<PhyPt>:CORRection:
SLEVeling:POWer:STEP is ignored.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<Compatible>
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "ALC Step Size" on page 604
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer:STARt <Leveling
Tool Sart Power>
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:STOP <PowValue>
Defines the start and stop power of the leveling grid for physical port <PhyPt>.