GUI reference
R&S
®
ZNA
603User Manual 1178.6462.02 ─ 29
Start Frequency/Stop Frequeny
Defines the start and stop frequency of the leveling grid.
By default the vector network analyzer uses the standard frequency range of the
selected "Meas Port".
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STARt
SOURce:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:STOP
Number of Points
Defines the number of frequency sweep points of the leveling grid.
By default the R&S ZNA uses 201 sweep points. The higher the number of "Sweep
Points", the more accurate the leveling, but also the longer it takes to create the level-
ing table). The number of sweep points is limited to 100001.
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling:FREQuency:POINts
Start Power/Stop Power
Defines the start and stop (source) power of the leveling grid.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer:STARt
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:STOP
Power Step Size
Defines the (source) power step size of the leveling grid..
Cal softtool