Command reference
R&S
®
ZNA
1285User Manual 1178.6462.02 ─ 29
<CalGroupFile> Optional string parameter to specify the name of the cal group
file. Cal group files must have the extension *.cal. The direc-
tory must not be specified; the analyzer always uses the default
path C:\Users\Public\Documents\Rohde-Schwarz\ZNA
\Calibration\Data. If there is no link between <Ch> and the
specified file, the command has no effect. If no file is specified,
the command resolves any link between <Ch> and an arbitrary
cal group file.
Example: See MMEMory:LOAD:CORRection
Usage: Setting only
Manual operation: See "Resolve Pool Link / Remove Pool Link" on page 640
MMEMory:LOAD:CORRection:SLEVeling<PhyPt>
<File>
Loads a leveling dataset for port <PhyPt> from the specified *.lev file (R&S ZNA-pro-
prietary, binary file format).
*.lev files can be created using MMEMory:STORe:CORRection:
SLEVeling<PhyPt>.
Suffix:
<PhyPt>
.
Physical port number
Setting parameters:
<File> String parameter containing the path and file name of the *.lev
file. If the path is omitted, the current directory is used (see
MMEMory:CDIRectory).
Usage: Setting only
Options: R&S ZNA-K8
Manual operation: See "Load/Save" on page 908
MMEMory:LOAD:CORRection:TCOefficient<Ch> <TraceFile>[, <Trace>[,
<DeEmbedding Position>]]
Loads two-port transmission coefficients from the specified power meter correction file
or trace file to channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)