Command reference
R&S
®
ZNA
1303User Manual 1178.6462.02 ─ 29
<CalGroupFile> String parameter to specify the name of the created cal group
file. There is no need to specify the directory path and file exten-
sion; the analyzer uses the default cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\ZNA
\Calibration\Data and a *.cal extension.
Example: See MMEMory:LOAD:CORRection
Usage: Setting only
Manual operation: See "Add / Add All... / Replace / Apply / Apply to All"
on page 639
MMEMory:STORe:CORRection:SLEVeling<PhyPt>
<File>
Persists the current leveling dataset of port <PhyPt> to the specified *.lev file
(R&S ZNA-proprietary, binary file format).
The dataset can be recalled using MMEMory:LOAD:CORRection:
SLEVeling<PhyPt>.
Suffix:
<PhyPt>
.
Physical port number
Setting parameters:
<File> String parameter containing the path and file name of the *.lev
file. If the path is omitted, the current directory is used (see
MMEMory:CDIRectory).
Usage: Setting only
Options: R&S ZNA-K8
Manual operation: See "Load/Save" on page 908
MMEMory:STORe:CORRection:TCOefficient<Ch> <PmclFile>
Saves two-port transmission coefficients of channel <Ch> to a power meter correction
list file.
Suffix:
<Ch>
.
Channel number
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
SCPI command reference