Command reference
R&S
®
ZNA
1328User Manual 1178.6462.02 ─ 29
Parameters for setting and query:
<DllName> Name of a loaded DLL (see [SENSe:]CDLL:LIST?)
Lowercase string, without dll extension
Manual operation: See "Task Type" on page 403
[SENSe:]CDLL:REMove <DllName>
Tells the analyzer firmware to remove a custom (external) DLL that was previously
added using [SENSe:]CDLL:ADD.
Note that adding/removing custom DLLs is only possible if the firmware is run as
administrator.
Setting parameters:
<DllName> Name of a loaded DLL (see [SENSe:]CDLL:LIST?)
Lowercase string, without dll extension
Usage: Setting only
Manual operation: See "Add / Remove" on page 404
[SENSe<Ch>:]CDLL[:STATe] <DllName>, <Boolean>
[SENSe<Ch>:]CDLL[:STATe]? <DllName>
Defines the state of (external) DLL <DllName>.dll in channel <Ch>.
An active DLL can provide traces or operate in permanent mode (see [SENSe<Ch>:
]CDLL:PERManent[:STATe]).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
DLL is active
OFF (0)
DLL is inactive
Parameters for setting and query:
<DllName> Name of a loaded DLL (see [SENSe:]CDLL:LIST?)
Lowercase string, without dll extension
Manual operation: See "Active" on page 403
[SENSe<Ch>:]CDLL:TASK:ADDRess <DllName>, <Task>, <Address>
[SENSe<Ch>:]CDLL:TASK:ADDRess? <DllName>, <Task>
Defines the address of an external device the external DLL <DllName> requires to
complete task type <Task> in channel <Ch>.
Suffix:
<Ch>
.
Channel number