Use [SENSe<Ch>:]CDLL:PERManent:TASK to select the type of task to be run.
Suffix:
<Ch>
.
Channel number
Parameters:
<Additional Input> String value, with DLL-specific syntax and semantics. See the
docs of your DLL for details.
Parameters for setting and query:
<DllName> Name of a loaded DLL (see [SENSe:]CDLL:LIST?)
Lowercase string, without dll extension
Manual operation: See "Additional Input" on page 403
[SENSe<Ch>:]CDLL:PERManent[:STATe] <DllName>, <Boolean>
[SENSe<Ch>:]CDLL:PERManent[:STATe]? <DllName>
Defines whether DLL <DllName> shall operate in permanent mode in channel <Ch>.
Use [SENSe<Ch>:]CDLL:PERManent:TASK and [SENSe<Ch>:]CDLL:
PERManent:ADDitional to configure the permanent task to be run.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Permanent mode enabled
OFF (0)
Permanent mode disabled
Parameters for setting and query:
<DllName> Name of a loaded DLL (see [SENSe:]CDLL:LIST?)
Lowercase string, without dll extension
Manual operation: See "Permanent" on page 403
[SENSe<Ch>:]CDLL:PERManent:TASK <DllName>, <Task>
[SENSe<Ch>:]CDLL:PERManent:TASK? <DllName>
Defines the type of task to be run, if custom DLL <DllName> is set to permanent mode
in channel <Ch> ([SENSe<Ch>:]CDLL:PERManent[:STATe] '<'DllName>',
ON).
Use [SENSe<Ch>:]CDLL:PERManent:ADDitional to further specify the task.
Suffix:
<Ch>
.
Channel number
Parameters:
<Task> Name of the task to be run permanently.