GUI reference
R&S
®
ZNA
532User Manual 1178.6462.02 ─ 29
Cal Premeas On
Activates/deactivates power correction from previous measurements for the current
channel.
Remote command:
SOURce<Ch>:POWer:CPRemeas:CSTate
Cal Premeas On (table area)
Enables or disables power correction from previous measurements for the respective
port.
Only takes effect, if Cal Premeas On is active for the current channel.
Remote command:
SOURce<Ch>:POWer<Pt>:CPRemeas[:STATe]
Meas Port/Meas Receiver (table area)
Defines the received signal to be used for power correction from previous measure-
ments, i.e. whose power the R&S ZNA tries to keep at the level defined in the port con-
figuration.
Remote command:
SOURce<Ch>:POWer<Pt>:CPRemeas:RECeiver
Max Iterations/Tolerance/Convergence (table area)
Defines the parameters for Premeas Mode "Once".
In general, cal premeas uses the convergence logic of a flatness cal. See "Flatness
Cal – Max Iterations" on page 628, "Flatness Cal – Tolerance" on page 628, "Flat-
ness Cal – Convergence" on page 628.
However, to prevent the generators from oscillating, for cal premeas convergence fac-
tors >1 are not allowed.
Stimulus softtool