POWer:Z<Ptout><Ptin>
Matched-circuit impedance (converted Z-parameter)
with output and input port number of the DUT, e.g.
Z
11
, Z
21
.
POWer:Y<Ptout><Ptin>
Matched-circuit admittance (converted Y-parameter)
with output and input port number of the DUT, e.g.
Y
11
, Y
21
.
POWer:KFACtor
POWer:MUFactor<Lev>
Stability factor K
Stability factors μ
1
or μ
2
VOLTage[:DC] DCIN1 | DCIN2
Output: DC 1 V, DC 10 V
DC Input 1 or 2
Note: The mnemonics POWer: and VOLTage: are not used in output strings.
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire] <Sensor>
Initiates a source power calibration for the source port <PhyPt> using an external
power meter no. 1 or 2. To initiate a source power calibration for arbitrary power
meters, use the alternative commands listed in the program example below.
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Setting parameters:
<Sensor> ASENsor | BSENsor
ASENsor external power meter Pmtr 1.
BSENsor external power meter Pmtr 2.
*RST: ASENsor
Example:
SOUR:POW3:CORR:COLL BSEN
Perform a source power calibration for port 3 using power meter
no. 2.
SOUR:POW:CORR:PMET:ID 2
Select power meter no. 2.
SOUR:POW:CORR:ACQ PORT,3
Perform a source power calibration for port 3 using the previ-
ously selected power meter no. 2.
Usage: Setting only
Manual operation: See "Start Cal Sweep" on page 600
TRACe:CLEar <MemTrace>
Deletes one of the memory traces Mem<n>[Trc<m>], where n = 1, ... 8.
R&S ZVR/ZVABT compatible commands