Command reference
R&S
®
ZNA
1720User Manual 1178.6462.02 ─ 29
Setting parameters:
<MemTrace> MDATa1 | MDATa2 | MDATa3 | MDATa4 | MDATa5 | MDATa6 |
MDATa7 | MDATa8
Identifier for the memory trace; see Table 7-18.
Range: MDATA<n> where <n> = 1 to 8.
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace
the active trace of channel 1 (omitted optional mnemonic
SENSe1).
TRAC:COPY "Mem_Pt20",CH1DATA
Copy the current state of the created trace to a memory trace
named "Mem_Pt20". The memory trace is not displayed.
DISP:WIND:TRAC2:FEED "MEM_PT20"
Display the created memory trace in the active diagram area
(diagram area no. 1).
Usage: Setting only
TRACe[:DATA][:RESPonse][:ALL]? <Response>
Returns the response values of the active data trace or memory trace (see Table 7-18).
Note: To read the response values of an arbitrary data or memory trace, use
CALCulate<Chn>:DATA. To read the response values of a trace acquired in single
sweep mode (INITiate<Ch>:CONTinuous OFF), use CALCulate<Chn>:DATA:
NSWeep:FIRSt?.
Query parameters:
<Response> CH1Data | CH2Data | CH3Data | CH4Data | CH1Mem |
CH2Mem | CH3Mem | CH4Mem | MDATa1 | MDATa2 | MDATa3 |
MDATa4 | MDATa5 | MDATa6 | MDATa7 | MDATa8
Response data of the selected trace, see Table 7-18.
The data is transferred in the data format defined via FORMat[:
DATA] and FORMat:DEXPort:SOURce. The unit is the default
unit of the measured parameter; see CALCulate<Ch>:
PARameter:SDEFine.
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace
the active trace of channel 1 (omitted optional mnemonic
SENSe1).
CALC:FORM MLIN; :FORM ASCII; FORM:DEXP:SOUR
FDAT
Select the trace data format: linear magnitude values, ASCII for-
mat and formatted trace data (1 value per sweep point).
TRAC? CH1DATA
Query the 20 response values of the created trace according to
the previous format settings.
Usage:
Query only
R&S ZVR/ZVABT compatible commands