Command reference
R&S
®
ZNA
1552User Manual 1178.6462.02 ─ 29
RRAFter
First the reference receiver will be calibrated by a single power
meter measurement, then the source flatness calibration will be
performed using the calibrated reference receiver.
Requires
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
FLATness ON.
RRONly
Reference receiver only. The reference receiver will be used to
perform a source flatness calibration. Ideally the reference
receiver was previously calibrated using a power meter.
Requires
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
RRECeiver OFF.
*RST:
RRAFter
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Source Leveling/Src. Lev. Method" on page 584
SOURce:POWer:CORRection:COLLect:RRECeiver <Boolean>
Enables or disables calibration of the reference receiver together with the source
power calibration.
Note: It is not possible to disable flatness calibration (see SOURce:POWer:
CORRection:COLLect:FLATness on page 1551) and reference receiver calibration
simultaneously.
Parameters:
<Boolean> Disables (OFF|1) or enables (ON|1) reference receiver calibra-
tion.
*RST: ON|1
Example: See SOURce:POWer:CORRection:COLLect:FLATness
on page 1551.
SOURce:POWer:CORRection:COLLect:VNETworks <DeEmbNwType>[,
<DeEmbType>]
Defines whether the source flatness cal uses the configured offset de-/embedding to
get the power values at the DUT (instead of the calibrated reference plane).
This setting applies to all physical ports and all channels.
SCPI command reference