GUI reference
R&S
®
ZNA
683User Manual 1178.6462.02 ─ 29
tions between physical VNA ports and DUT ports. Furthermore it provides convenient
access to port-related settings.
New in firmware version 2.30
On a R&S ZNA with four internal sources, four 1-port DUTs can be measured in paral-
lel.
Related information
See Chapter 4.1.4.2, "Parallel measurements on multiple DUTs", on page 112.
The "Parallel Measurement" is automatically configured when "Enabled" is selected, as
shown in Figure 5-35.
A R&S ZNA with two internal sources allows parallel measurements with "Frequency
Offset", which can be configured in the corresponding tab on the right.
Figure 5-35: Define Parallel Measurement
●
Modified settings are not applied unless the "Define Parallel Measurement" dialog
is closed with the "OK" button.
●
On "OK", any pre-existing logical port configuration is overwritten.
●
Each configured DUT can have its own ground loop embedding and deembedding
network
Channel Config softtool