Command reference
R&S
®
ZNA
1530User Manual 1178.6462.02 ─ 29
Parameters:
<Power> Power of the CW signal
*RST: -10 dBm
Default unit: dBm
Manual operation: See "Frequency / Power" on page 921
SOURce:INDependent:CLO[:STATe] <Active>
Defines whether the converter LO Out port operates as independent CW source.
Parameters:
<Active> ON (1)
The converter LO port operates as independent CW source.
OFF (0)
The converter LO port can be used for other (channel-related)
purposes.
*RST: OFF
Options: R&S ZNA-B8
Manual operation: See "Independent Gen" on page 921
SOURce:INDependent<Pt>:ATTenuator <Step Attenuator>
If RF port <Pt> is configured as an independent CW source (SOURce:
INDependent<Pt>[:STATe] ON) and equipped with a source step attenuator, this
command sets the source attenuation to be applied.
Suffix:
<Pt>
.
VNA RF port number
Parameters:
<Step Attenuator> Source step attenuation
Range: 0 dB to 70 dB
Increment: 10 dB
*RST: 0 dB
Default unit: dB
Options: R&S ZNAxx-B2<Pt>
Manual operation: See "Step Att" on page 921
SOURce:INDependent<Pt>:FREQuency <Frequency>
If RF port <Pt> is configured as an independent CW source (SOURce:
INDependent<Pt>[:STATe] ON), this command sets the frequency of the CW sig-
nal.
Suffix:
<Pt>
.
VNA RF port number
SCPI command reference