Command reference
R&S
®
ZNA
1035User Manual 1178.6462.02 ─ 29
OFF (0)
The frequencies only reach down to the start frequency of the
instrument sweep.
Options: R&S ZNA-K210
Manual operation: See "DC Extrapolation" on page 742
CALCulate:FMODel:EZD:DUT:MEASure
For a fixture modeling with the EZD tool, this command allows to measure "DUT + Test
Fixture" at the active ports (see CALCulate:FMODel:EZD<Ph_pt>:DUT[:STATe])
Usage:
Event
Options: R&S ZNA-K210
Manual operation: See "Measure" on page 733
CALCulate:FMODel:EZD:DUT:MEASure:FILename <String>
For a fixture modeling with the EZD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
Options: R&S ZNA-K210
Manual operation: See "Load File" on page 733
CALCulate:FMODel:EZD:DUT:MEASure:FILename:CLEar
Invalidates the file path previously set using CALCulate:FMODel:EZD:DUT:
MEASure:FILename.
Usage:
Event
Options: R&S ZNA-K210
Manual operation: See "Load File / 1x Open Preset / 1x Short Preset" on page 732
CALCulate:FMODel:EZD<Ph_pt>:DUT[:STATe] <Boolean>
Defines the physical ports to which DUT + test fixture are connected.
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> ON (1)
Port <Ph_pt> connected
SCPI command reference