GUI reference
R&S
®
ZNA
828User Manual 1178.6462.02 ─ 29
Remote command:
CALCulate<Chn>:DTIMe:STATe
Reference Trace
Selects the reference trace for the skew measurement.
Note that the reference trace must be in the same channel as the active trace.
See Skew for additional conditions on both the active and the reference trace.
Remote command:
CALCulate<Chn>:DTIMe:TARGet
Skew Position
Defines the percentage of the step size that defines the position of the reference points
on the current and the reference trace. The "Skew" value is the delta between the X
coordinates of these reference points.
Remote command:
CALCulate<Chn>:DTIMe:POSition
Extended Info
Defines whether the info field for the skew measurement shall only provide basic infor-
mation (default) or extended information:
Figure 5-66: Skew info, basic vs. extended
The displayed skew values are positive, if, at the defined Skew Position, the active
trace is to the right of the Reference Trace.
Remote command:
CALCulate<Chn>:DTIMe:DATA?
Skew Check On
Enables the skew check with the specified Skew Limit.
Remote command:
CALCulate<Chn>:DTIMe:LIMit:STATe
Skew Limit
Defines the limit value for the skew check. The unit depends on the current format of
the X axis (see " Time / Distance" on page 535).
An acoustic signal and a TTL signal indicating pass or fail can be generated in addi-
tion.
Remote command:
CALCulate<Chn>:DTIMe:LIMit:LIMit
Skew Fail Beep
This checkbox determines whether the R&S ZNA should make an audible beep on
skew limit violations.
Applic softtool