Command reference
R&S
®
ZNA
984User Manual 1178.6462.02 ─ 29
Query parameters:
<Format> FDATa | SDATa | MDATa
Output format for the S-parameter data, see
CALCulate<Chn>:DATA on page 975.
Example: See CALCulate<Ch>:PARameter:DEFine:SGRoup
Usage: Query only
CALCulate<Chn>:DATA:STIMulus?
Reads the stimulus values of the active data or memory trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example: See CALCulate<Chn>:DATA
Usage: Query only
CALCulate:DATA:TRACe <TraceName>, <Format>, <Data>...
The query gets the trace data of an arbitrary (not necessarily the active) trace, refer-
enced by its trace name <TraceName>.
The "set direction" allows to import formatted or unformatted trace data to an existing
trace.
Note
●
Unformatted trace data (SDATa) can only be imported to memory traces
●
Formatted trace data (FDATa) can only be imported to "live" traces if the related
channel is in single sweep mode (INITiate<Ch>:CONTinuous OFF).
Before the import, the target trace must be prepared according to the settings used
during export. Any request for new data from the hardware ("Restart Sweep" in sin-
gle sweep mode or switching to continuous sweep mode) discards imported data
and switches back the display to measured data.
Parameters:
<TraceName> String parameter containing the trace name
<Format> FDATa | SDATa | MDATa | NCData | UCData
Data format; see Table 7-2.
<Data> Trace data for FDATa | SDATa import either in ASCII or block
data format, depending on the current FORMat[:DATA] setting.
The column order must match the one used during export.
Example: See CALCulate<Chn>:DATA.
SCPI command reference