Command reference
R&S
®
ZNA
1642User Manual 1178.6462.02 ─ 29
Parameters:
<Options> String parameter containing the available options
Example: See SYSTem:OPTions:FACTory
Manual operation: See "Define *IDN + *OPT..." on page 889
SYSTem:PASSword[:CENable] <Password>
Sends a password to the analyzer enabling a class of service functions to function
(Command ENable). Service functions are activated with the commands of the
DIAGnostic... system and should be used by a R&S service representative only.
Setting parameters:
<Password> Case-sensitive string variable. Sending an invalid password gen-
erates error -221, (settings conflict).
Example:
SYST:PASS "XXXX"
Enter password.
Usage: Setting only
Manual operation: See "Password" on page 886
SYSTem:PRESet[:DUMMy]
Performs a preset, whose scope is defined using the SYSTem:PRESet:SCOPe com-
mand. The SYSTem:PRESet:USER[:STATe] command determines whether a factory
preset or a user-defined preset is performed.
Example:
SYST:PRES:SCOP SING
Define the scope of a preset: the active recall set is reset only.
SYST:PRES
Reset the parameters of the current recall set to their factory
defaults.
SYSTem:PRESet:MODE <PresetMode>
Allows you to select between different preset modes.
GLOBal and RFOFf presets can only be performed from the analyzer GUI. FILE pre-
sets can also be performed via remote commands, if SYSTem:PRESet:REMote[:
STATe] is set to ON.
Parameters:
<PresetMode> NORMal | FILE | GLOBal | RFOFf
NORMal
Factory preset
FILE
User preset file, see SYSTem:PRESet:USER:NAME
on page 1644