Command reference
R&S
®
ZNA
1644User Manual 1178.6462.02 ─ 29
SYSTem:PRESet:STARt <Boolean>
Defines whether the analyzer firmware always starts with the configured system preset
(factory or user defined) or with the previous state. This is a global setting.
Parameters:
<Boolean> ON
The FW always starts with the configured system preset.
See SYSTem:PRESet:USER:NAME and SYSTem:PRESet:
USER[:STATe].
OFF
The analyzer firmware recalls the previous state, i.e. the settings
that were persisted during the previous shutdown.
Manual operation: See "Start in Preset" on page 858
SYSTem:PRESet:USER:CAL
<PresetUserCal>
Selects a calibration from the calibration pool that shall be restored during a user-
defined preset.
The corresponding cal group file (<PresetUserCal>.cal) must be available in the
cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\ZNA\Calibration\Data\.
Parameters:
<PresetUserCal> Name of the cal group (i.e. the name of the cal group file without
path and extension).
Use the empty string to restore the default behavior (no preset
user cal).
Example:
MMEMORY:STORE:CORRection 1, 'Refl1
Trans123.cal'
Adds the active user calibration to the memory pool
SYSTem:PRESet:USER:CAL 'Refl1 Trans123'
Sets the newly created cal pool member as the preset user cal.
SYSTem:PRESet:USER:CAL?
Queries for the preset user cal. Returns 'Refl1 Trans123'
SYSTem:PRESet:USER:CAL ''
Restores the default behaviour of the instrument: no preset user
cal.
Manual operation: See "Preset User Cal" on page 640
SYSTem:PRESet:USER:NAME <RecallSetFile>
Specifies the name of a recall set file (*.znxml|*.znx) to be used for a user-defined
preset.