Command reference
R&S
®
ZNA
1643User Manual 1178.6462.02 ─ 29
GLOBal
Factory preset + GUI reset + reset RF connection configuration
between VNA and external devices
RFOFf
Factory preset + OUTPut<Ch>[:STATe] OFF
Manual operation: See "Normal / RF Off / Normal, GUI, Ext Setup / <File Name>"
on page 929
SYSTem:PRESet:REMote[:STATe] <Boolean>
Defines the behavior of the preset commands *RST and SYSTem:PRESet[:DUMMy]
Parameters:
<Boolean> OFF (0)
*RST and SYSTem:PRESet perform a factory preset.
ON (1)
*RST and SYSTem:PRESet restore the user preset file selected
using SYSTem:PRESet:USER:NAME, if this file is available and
valid. Otherwise, they perform a factory preset.
Manual operation: See "Remote Preset Configuration" on page 857
SYSTem:PRESet:SCOPe <Scope>
Specifies whether the preset actions affect all open recall sets, or only the active recall
set.
Applies to the GUI and the SYSTem:PRESet[:DUMMy] command. The *RST com-
mand always resets all open recall sets.
Parameters:
<Scope> ALL | SINGle
ALL
All open recall sets are deleted and a new recall set"Set1" is cre-
ated according to the current preset mode.
See SYSTem:PRESet:MODE on page 1642.
SINGle
The active setup is reset according to the current preset mode;
the name of the active setup and the parameters of all other set-
ups remain unchanged.
Example: See SYSTem:PRESet[:DUMMy]
Manual operation: See "Preset Scope" on page 857