GUI reference
R&S
®
ZNA
857User Manual 1178.6462.02 ─ 29
Preset Scope
Defines whether a preset affects all open recall sets ("Instrument") or the active recall
set only.
The "Preset Scope" applies to the GUI and to the SYSTem:PRESet[:DUMMy] com-
mand. The *RST command always resets all open recall sets.
Remote command: 
SYSTem:PRESet:SCOPe
Remote Preset Configuration
"Align *RST to User Defined Preset" defines the behavior of the *RST and
SYSTem:PRESet commands. If checked, and if a valid setup file is selected in the Pre-
set Configuration section, then they perform a user-defined preset. Otherwise, they
perform a factory preset.
Note that the preset modes "RF Off" and "Normal, GUI, Ext Setup" are not available via
remote commands.
Remote command: 
SYSTem:PRESet:REMote[:STATe]
Global Settings
The two buttons reset all directory settings (e.g. the directories for storing trace data,
limit lines, calibration data...) and all settings in the "Printer Setup" dialog to default val-
ues. See Chapter 5.15.4, "Printer Setup dialog", on page 779.
Remote command: 
n/a
Preset Configuration
Specifies whether the preset action performs a factory preset or restores a user preset
file. A user preset file is an arbitrary recall set file (*.znxml|*.znx), that can be cre-
ated from the active setup using System – [File Print] > "Save...". If the current user
preset file is not found (e.g. because it was deleted or moved), System – [Preset] ini-
tiates a factory preset.
Setup softtool