Command reference
R&S
®
ZNA
1229User Manual 1178.6462.02 ─ 29
CONTrol:VSE:DISConnect
Disconnects an active R&S VSE connection (see CONTrol:VSE:CONNect)
Usage: Event
Manual operation: See "Start/Connect VSE / Disconnect VSE" on page 701
CONTrol:VSE:LOCal <Boolean>
Tells the R&S ZNA whether you want to connect to a local or remote R&S VSE
instance.
Use CONTrol:VSE:CONNect on page 1228 to establish the connection.
Parameters:
<Boolean> 1
Local R&S VSE
0
Remote R&S VSE
*RST: 0
Manual operation: See "Local VSE" on page 701
CONTrol:VSE:MODE <VseMode>
Determines where the acquired measurement data are processed.
Parameters:
<VseMode> DEVice | VSE | BOTH
DEVice
Only on the R&S ZNA
VSE
Only on the R&S VSE
BOTH
On the R&S ZNA and on the R&S VSE
Manual operation: See "VSE Mode" on page 692
CONTrol:VSE:TRACe <No>
Select the trace to be pushed to R&S VSE, if connected (see CONTrol:VSE:
CONNect) and enabled (see CONTrol:VSE:MODE).
Parameters:
<No> Trace number
Manual operation: See "Trace (b-wave)" on page 693
SCPI command reference