104User Manual 1178.6462.02 ─ 29
b) "Meas. Receiver"
A measurement receiver calibration adjusts the power readings at the receive
port, by default based on an existing reference receiver calibration.
With an existing reference receiver calibration for port 1, to calibrate the mea-
surement receiver of port 1, connect an Open to the 1 mm port of converter 1
and select port 1 as source.
To calibrate the measurement receiver of port 2, connect the 1 mm ports of the
test sets directly, and again use port 1 as source.
Getting started with R&S ZNA67EXT