Command reference
R&S
®
ZNA
1061User Manual 1178.6462.02 ─ 29
Parameters:
<CenterX> X position (real part) of the display circle's center
<CenterY> Y position (imaginary part) of the display circle's center
<Radius> Radius of the display circle
Manual operation: See "Draw Circle / Radius, Center X, Center Y" on page 497
CALCulate<Chn>:LIMit:DCIRcle:DISPlay[:STATe] <Boolean>
Sets/queries the visibility of the display circle for the active trace of channel <Chn>.
The display circle is defined using CALCulate<Chn>:LIMit:DCIRcle:DATA.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> If set to ON, the line of the display circle is shown whenever the
related trace is displayed in complex format.
Manual operation: See "Show Border" on page 496
CALCulate<Chn>:LIMit:DCIRcle[:STATe] <Boolean>
Sets/queries the state of the display circle for the active trace of channel <Chn>.
The display circle is defined using CALCulate<Chn>:LIMit:DCIRcle:DATA.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> If set to ON, only trace points within the display circle are shown
at the GUI whenever the related trace is displayed in complex
format.
Manual operation: See "Limit to Circle On/Off" on page 496