EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1069

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1069User Manual 1178.6462.02 ─ 29
Example:
*RST
SENS:FREQ:STAR 1GHz
SENS:FREQ:STOP 3GHz
CALC:LIM:CONT 1GHz, 2GHz, 2GHz, 3GHz
CALC:LIM:SEGM1:FORM '22-(20/25.78)*StimVal/1e9'
CALC:LIM:SEGM1:FORM:STAT ON
CALC:LIM:SEGM2:FORM '15-(6/25.78)*StimVal/1e9'
CALC:LIM:SEGM2:TYPE LMIN
CALC:LIM:DISP ON
CALC:LIM:STAT ON
Manual operation: See "Linear/Formula" on page 486
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula:STATe
<State>
Defines whether a custom formula is used for limit line segment <Seg> of the related
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
If you want to use a custom formula for a segment, use
CALCulate<Chn>:LIMit:CONTrol[:DATA] to create it.
Parameters:
<State> OFF (0)
The line segment is a straight line, connecting the endpoints
specified in the segment definition.
ON (1)
The line segment is defined by a custom formula. Use
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula to spec-
ify the formula.
*RST: OFF
Example: See CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula
on page 1068
Manual operation: See "Linear/Formula" on page 486
CALCulate<Chn>:LIMit:SEGMent<Seg>:INTerpol <LimLineInterpolation>
"Interpolation" determines whether the limit line segment is interpolated linearly or log-
arithmically. The latter may be more suitable for logarithmic sweeps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
Parameters:
<LimLineInterpolation>LOGarithmic | LINear
SCPI command reference

Table of Contents

Related product manuals