Command reference
R&S
®
ZNA
1071User Manual 1178.6462.02 ─ 29
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
Parameters:
<LimLineType> LMIN | LMAX | OFF
Limit line type
Range: LMAX (upper limit line segment), LMIN (lower limit
line segment), OFF (limit check switched off, limit
line segment not deleted)
*RST: LMAX
Example:
*RST; :CALC:LIM:UPP 0, 0
Define an upper limit line segment across the entire sweep
range, using a constant upper limit of 0 dBm.
CALC:LIM:SEGM:TYPE LMIN
Turn the defined limit line segment into a lower limit line seg-
ment.
Manual operation: See "Segment List" on page 483
CALCulate<Chn>:LIMit:SOUNd[:STATe] <Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Fail beep on or off.
*RST: OFF
Example:
CALC:LIM:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
Manual operation: See "Limit Fail Beep" on page 480
CALCulate<Chn>:LIMit:STATe <Boolean>
Switches the limit check (including upper and lower limits) on or off.
Tip: Use CALCulate<Ch>:LIMit:UPPer:STATe or
CALCulate<Ch>:LIMit:LOWer:STATe to switch on or off the individual limit checks
for upper or lower limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Limit check on or off.
*RST: OFF
SCPI command reference