EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1085

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1085User Manual 1178.6462.02 ─ 29
CALCulate<Chn>:MARKer<Mk>:FUNCtion:SPAN
Sets the span of the sweep range equal to the absolute value of the first coordinate of
the active delta marker <Mk> on trace no. <Chn>.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number.
Example:
*RST; :CALC:MARK ON; MARK:DELTa ON
Create marker 1 in the center of the current sweep range and
enable the delta mode.
CALC:MARK:X 300MHz
Increase the stimulus value of the delta marker by 300 MHz.
CALC:MARK:FUNC:SPAN
Set the sweep range equal to 300 MHz. The sweep range starts
at the reference marker position, i.e. in the center of the analyz-
er's frequency range.
Usage: Event
Manual operation: See "Center = Marker / Start = Marker / Stop = Marker / Span =
Marker" on page 519
CALCulate<Chn>:MARKer<Mk>:FUNCtion:SPECtrum
For marker <Mk> of channel <Chn>'s active trace, this command does the following:
Create a channel
Set its center to marker <Mk>'s position
Set its span to 10 times channel <Chn>'s BW
Switch to spectrum mode
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number
Usage: Event
Options: R&S ZNA-K1
Manual operation: See "Spectrum = Marker" on page 506
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:STOP
These commands set the STARt or STOP of the sweep range to the stimulus value of
the related marker (see "Marker addressing" on page 1073).
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference

Table of Contents

Related product manuals