EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1090

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1090User Manual 1178.6462.02 ─ 29
Parameters:
<SearchFormat> MLINear | MLOGarithmic | PHASe | UPHase | REAL |
IMAGinary | SWR | DEFault
The format of the target value in the marker target search
(CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute TAR
Get or CALCulate<Chn>:MARKer<Mk>:REFerence:
FUNCtion:EXECute TARGet). See table above.
*RST: DEFault
Example: Suppose channel 1's selected trace is POLar and marker 1 isn't
yet created
:CALCULATE1:MARKER1 ON
Create/enable Marker 1
:CALCulate1:MARKer1:FUNCtion:SELect TARGet
Select TARGet search mode for marker 1
:CALCulate1:MARKer1:SEARch:FORMat?
Query the target format of marker 1. The result is DEF and for
polar diagrams the default target format is "Phase".
:CALCulate1:MARKer1:FUNCtion:TARGet?
Query for the default target value; for "Phase" this is 0 (degrees)
:CALCulate1:MARKer1:SEARch:FORMat MLOGarithmic
Change the target search format to dB magnitude
:CALCulate1:MARKer1:FUNCtion:TARGet?
Query for the default target value; for dB magnitude this is 0 (dB)
:CALCulate1:MARKer1:FUNCtion:TARGet -3
Set the target value to -3 dB
:CALCulate1:MARKer1:FUNCtion:EXECute
Execute the target search for marker 1
:CALCulate1:MARKer1:FUNCtion:RESult?
Query for the results.
Manual operation:
See "Target Format" on page 513
CALCulate<Chn>:MARKer<Mk>:SEARch:TRACking <Boolean>
CALCulate<Chn>:MARKer<Mk>:REFerence:SEARch:TRACking <Boolean>
These commands enable or disable tracking for the respective marker (see "Marker
addressing" on page 1073).
They are only available, if a search mode is active for the respective marker (see
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute <SearchMode> or
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute <Search
Mode>), in which case tracking causes the search logic to be repeated after each
sweep.
Tip: If the current search mode is a bandfilter or multiple peak search (regular markers
only), this command enables or disables the corresponding tracking.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference

Table of Contents

Related product manuals