Command reference
R&S
®
ZNA
1162User Manual 1178.6462.02 ─ 29
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
TNDefinition <CircuitModel>
Selects the circuit model for balanced port embedding.
Suffix:
<Ch>
.
Channel number
<LogPt> Logical port number (balanced port)
Parameters:
<CircuitModel> FIMPort | STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC |
LSSL | SLCS | SCLS | SCCS | SLLS | STSG | SGST | GSSL |
LSSG | GSSG | SLGS | SGLS | SGGS
Possible circuit models (character data); see Table 7-7.
*RST: CSSL
Example: See CALCulate<Ch>:TRANsform:VNETworks:BALanced:
EMBedding<LogPt>[:STATe]
Manual operation: See "Network" on page 747
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:C<Cmp> <CircuitModel>[, <Capacitance>]
Specifies the capacitance value C in the "Shunt L, Shunt C" lumped element model for
differential match embedding.
In the query form, the <Capacitance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<LogPt> Logical port number (balanced port)
<Cmp> Must be omitted or set to 1.
Parameters:
<CircuitModel> SHLC
Currently only the "Shunt L, Shunt C" lumped element model is
supported
<Capacitance> Range: -1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST: 1 pF (1E-12 F)
Default unit: F
Manual operation: See "Network" on page 752
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:DATA <Interchange>, <SParamTrcs>
Defines a Differential match embedding network for a balanced port based on the
given S-Parameter traces.