EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1185

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1185User Manual 1178.6462.02 ─ 29
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Setting parameters:
<Interchange> FPORts | IPORts | SGATes
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
<arg1> <block_data>
Content of a Touchstone file (*.s2p) in IEEE488.2 Block data
format.
Usage: Setting only
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp> <CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>? <CircuitModel>
Specifies the conductance value G in the "shunt L, shunt C" circuit model for single
ended port deembedding.
In the query form, the <Conductance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
<Cmp> Number of the conductance component in the circuit model. The
total number of conductances depends on the selected circuit
model.
Parameters:
<Conductance> Conductance G<Cmp> for the specified circuit model.
Range: -1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST: 0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel> SHLC
Circuit model whose conductance G<Cmp> shall be set, see
Table 7-6.
SCPI command reference

Table of Contents

Related product manuals