EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 1249

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1249User Manual 1178.6462.02 ─ 29
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
Manual operation: See "Add Trace" on page 432
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:LABel:SHOW <TraceName>[,
<Boolean>]
Displays or hides the label of an existing trace, identified by its <TraceName>.
Suffix:
<Wnd>
.
ignored
<WndTr> ignored
Parameters:
<TraceName> Trace name (string parameter), e.g. 'Trc4'
<Boolean> ON | OFF display or hide the label of the related trace
Example:
*RST; :DISP:TRAC:LAB:SHOW? 'Trc1'
Reset the analyzer, creating the default trace 'Trc1'. The trace
label is displayed; the query returns 1.
DISP:TRAC:LAB:SHOW 'Trc1', OFF
Hide the label of trace 'Trc1'.
Manual operation: See "Table Area" on page 435
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:SHOW <TraceName>[, <Boolean>]
Displays or hides an existing trace, identified by its trace name <Trace_Name>, or a
group of traces.
Tip: You can open the trace manager to obtain an overview of all channels and traces,
including the traces that are not displayed.
Suffix:
<Wnd>
.
Number of a diagram area. This suffix is ignored; the command
affects traces in all diagram areas.
<WndTr> Trace number. This suffix is ignored; the trace is referenced by
its name.
Parameters:
<TraceName> DALL all data traces
MALL all memory traces
<string> single trace identified by its trace name (string param-
eter), e.g. 'Trc4'.
SCPI command reference