Command reference
R&S
®
ZNA
1289User Manual 1178.6462.02 ─ 29
Example: Assume that the current recall set contains two traces named
Trc1 and Trc2, respectively, and that limit lines have been
defined for Trc1.
MMEM:STOR:LIM 'TRC1', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\LIMitLines\Lim_Trc1.limit'
Store the limit line definition of Trc1 to a limit line file.
MMEM:LOAD:LIM 'TRC2', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\LIMitLines\Lim_Trc1.limit'
Load the previously created limit line file and assign the limit
lines to Trc2.
MMEM:STOR:TRAC 'TRC1', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\LIMitLines\Trc1.s1p'
Store the current trace data of Trc1 to a limit line file in Touch-
stone format.
MMEM:LOAD:LIM 'TRC1', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\LIMitLines\Trc1.s1p', 'S11',
0, 2, LMAX
Load the previously created Touchstone limit line file and assign
the limit lines to Trc1, applying a response offset of 2 dB.
CALC:LIMit:DISPlay ON
Show the limit line in the diagram.
Usage:
Setting only
Manual operation: See "Recall... / Save..." on page 484
MMEMory:LOAD:MDAData <Channel Id>, <Filename or CDELAy>
Loads the known delay values of a calibration mixer, to be used as a reference for a
mixer delay measurement calibration. Mixer delay measurements are controlled using
the SENSe<Ch>:FREQuency:MDELay... commands.
Setting parameters:
<Channel Id> Channel number
<Filename or
CDELAy>
String parameter to specify the name and directory of the loaded
file. The default extension (manual control) for files containing
known delays is *.csv, although other extensions are allowed.
If no path is specified the analyzer searches the current direc-
tory, to be queried with MMEMory:CDIRectory?.
Example: See MMEMory:LOAD:MDCData on page 1290
Usage: Setting only
Manual operation: See "Variable Cal Delay File" on page 592
SCPI command reference