Command reference
R&S
®
ZNA
1351User Manual 1178.6462.02 ─ 29
Example:
CORR:CKIT:FOP:WLAB 'N 50 Ohm','New Kit 1',
'Test kit created today', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Create a new cal kit "New Kit 1" labelled "Test kit created today"
and assign an open (f) standard for the N 50 Ω connector type
with specific properties.
CORR:CKIT:LLAB 'New Kit 1', 'Test kit created
today', '2012-05-25'
Change the label of the previously created kit.
CORR:CKIT:LLAB? 'New Kit 1', '2012-05-25'
Check the label.
MMEMory:STORe:CKIT:
WLABel 'New Kit 1', '2012-05-25', 'C:
\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\New Kit 1
(2012-05-25).calkit'
Store the data for the labelled cal kit to the cal kit file
New Kit 1 (2012-05-25).calkit.
CORR:CKIT:LDEL 'New Kit 1', '2012-05-25'
Delete the kit. from the internal memory.
MMEMory:LOAD:CKIT 'C:\Users\Public\Documents
\Rohde-Schwarz\ZNA\Calibration\Kits\New Kit 1
(2012-05-25).calkit'
Re-load the kit.
[SENSe:]CORRection:CKIT:LSELect <ConnectorType>, <CalKitName>, <KitLabel>
Selects the calibration kit to be used, specifying its connector type, name, and label.
Tip: The command is suitable for connector types with arbitrary, user-defined names.
For standard connector types you can use the command [SENSe:]CORRection:
CKIT:<ConnType>:LSELect .
Parameters:
<ConnectorType> Connector type, e.g. a user-defined connector type (string vari-
able).
<CalKitName> String parameter containing the name of a calibration kit availa-
ble on the analyzer.
<KitLabel> String parameter containing the label of a calibration kit available
on the analyzer.
SCPI command reference