GUI reference
R&S
®
ZNA
351User Manual 1178.6462.02 ─ 29
This dialog can only be opened, if a balanced and two single-ended logical ports are
configured in the Balanced Ports dialog.
The Logical Port on the left represents the balanced test port: any (active) bal-
anced logical port can be selected
Physical Port 1 and Physical Port 2 represent the two single-ended test ports: any
two (active) single-ended logical ports can be selected
The part below allows you to select the signal direction and hence the "Imbalance
Result" to be calculated.
DUT
Balanced
port
Balanced
port
Log.
NWA
port
Physical port c
Logical ports j,k
(single-ended)
Physical port d
Logical port i
(balanced)
Physical port a
Physical port b
Imbalance 1 / Imbalance 2
Imbalance 1 selects the balanced logical port i as the input, the single-ended logical
ports j and k as the output and calculates the following imbalance parameter:
Imb
jk-i
= –(S
ca
– S
da
)/(S
cb
– S
db
)
Imbalance 2 selects the balanced logical port i as the output, the single-ended logical
ports j and k as the input and calculates the following imbalance parameter:
Imb
i-jk
= –(S
ac
– S
ad
)/(S
bc
– S
bd
)
Meas softtool