EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 479

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
479User Manual 1178.6462.02 ─ 29
The limit line colors are defined in the Define User Color Scheme dialog (System >
[Display] > "Config" > "Define User Color..."). You can choose between various options:
Display upper and lower limit lines with different colors.
Assign the same color to traces and associated limit lines.
Assign different colors to limit line segments with disabled limit check.
Note: Displaying the limit line and limit check are independent of each other: Hiding
the limit line does not switch off the limit check.
Remote command:
CALCulate<Chn>:LIMit:DISPlay[:STATe]
Limit Check
Enables/disables the limit check for the active trace. If enabled, an additional info field
is displayed in the diagram, indicating the "PASS" or "FAIL" state.
Limit violations are marked with squares, whose color is defined in the Define User
Color Scheme dialog ("Limit Fail Trace Color").
An acoustic signal and a TTL signal indicating pass or fail can be generated in addi-
tion.
Lines softtool

Table of Contents

Related product manuals