GUI reference
R&S
®
ZNA
551User Manual 1178.6462.02 ─ 29
Pulse Profile
This sweep type is only available, if the R&S ZNA is equipped with software option
R&S ZNA-K7 Pulsed measurements.
Activates the pulse profile sweep defined in the Pulse Modulation dialog.
Remote command:
[SENSe<Ch>:]SWEep:TYPE PULSe
Phase Imbalance/ Amplitude Imbalance
These sweep types are only available, if the R&S ZNA is equipped with software option
R&S ZNA-K6 Phase coherent source control. They can only be activated if source
coherence or True Differential Mode is active.
Remote command:
[SENSe<Ch>:]SWEep:TYPEIPHase | IAMPlitude
Define Segments...
Opens the Define Segments dialog that allows to set up the channel for a Segmented
frequency sweep.
Seg X-Axis
Scales the x-axis for a segmented frequency sweep:
In "Freq based" mode, the x-axis covers the frequency ranges of all sweep seg-
ments, including possible gaps between the segments. Equal frequency spacings
correspond to equal distances on the x-axis.
In "Point based" mode, the x-axis shows all sweep points with equal spacings.
Gaps between sweep segments are minimized; no diagram space is "wasted" on
unused frequency ranges. Point-based mode is indicated in the channel line.
The example below shows a segmented frequency sweep with two segments. The first
segment ranges from 1 GHz to 1.4 GHz; the second segment from 2 GHz to 3 GHz.
Both segments contain 11 sweep points. In point-based mode (lower diagram), all
sweep points are equidistant.
Sweep Softtool