GUI reference
R&S
®
ZNA
783User Manual 1178.6462.02 ─ 29
The "DUT Centric Wizard" lets you proceed with the followng steps:
1. Select the "DUT Type".
Currently, this can be either "Mixer" or "Amplifier", where "Mixer" measurements
require option R&S ZNA-K4 (see Chapter 4.7.3, "Frequency conversion measure-
ments", on page 261). Other DUT types may be added in future releases of the
analyzer firmware.
Currently the wizard only offers support for "Amplifier" measurements. Other DUT
types may be added in future releases of the analyzer firmware.
2. Specify the (basic) properties of your DUT.
3. Specify how the DUT is connected to the VNA.
4. Select the measurements to be performed.
At this point, the wizard knows how many additional channels are required, and
how they should be set up per default.
5. Adjust the default channel settings to your needs.
6. Review the channels to be created and the parameters to be measured. Decide
about calibration.
When you finish the "DUT Centric Wizard", the analyzer firmware creates and con-
figures the required channels for you. If you have decided to proceed with calibra-
tion, the wizards selects the suitable calibration types and guides you through the
channel calibrations (see Chapter 5.11.1.7, "Guided Calibration wizard",
on page 607).
DUT softtool

Table of Contents

Related product manuals