Power Reduction at Sweep End
The power reduction settings apply to all sweep modes but are particularly useful in
single sweep mode.
Keep
Auto
Reduce
Power
Time
Settling Delay
End Measurement Restart Sweep
Figure 5-70: Power control at sweep end in Single Sweep mode (power sweep)
Note: By default, at sweep end the output power of the first measurement point is
restored ("Auto" mode). After a small settling time (~10 ms), the sweep can be restar-
ted.
Selecting a different behavior can result in a significantly longer total measurement
time, in particular if a long "Settling Delay" is used. Do not forget to revert the Power
Mode at Sweep End to "Auto" mode if it is not necessary to "Reduce" or "Keep" the
power at the sweep end. A preset is not sufficient for this purpose.