EasyManua.ls Logo

Rohde & Schwarz R&S ZNA Series - Page 892

Rohde & Schwarz R&S ZNA Series
1849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
892User Manual 1178.6462.02 ─ 29
The configuration of a new external power meter involves the following steps:
1. Connect the power meter to your R&S ZNA using a LAN (VXI-11), GPIB, or USB
interface.
2. If the power meter is connected via LAN, enable LAN Detection
3. Select Scan Instruments and wait until the power meter appears in the table of
"Known Devices".
4.
Select to add the power meter to the list of Configured Devices.
If the R&S ZNA fails to detect a connected power meter:
Select Add Device to define the interface type and address.
The R&S ZNA can auto-detect the instrument type (driver) and the serial number of
the connected power meter.
To control external devices via GPIB, the pre-installed RS Visa library (visa32.dll)
must be replaced by the NI Visa library.
Known Devices
Table with all power sensors/meters the analyzer detects to be on line (i.e. connected
and switched on). "Scan Instruments" refreshes the table.
Select to add a power sensor/meter to the table of Configured Devices.
Note: Even though an R&S NRP2 power meter can appear in the table of "Known
Devices", adding it to the list of "Configured Devices" directly results in an invalid con-
figuration. Use Add Device to configure it manually instead.
Remote command:
n/a
Setup softtool

Table of Contents

Related product manuals