GUI reference
R&S
®
ZNA
924User Manual 1178.6462.02 ─ 29
LAN Detection
Activate "LAN Detection" to include the LAN interface in the autodetection sequence
(see Scan Instruments).
Note: "LAN detection" only works for external devices sharing the IP subnet with the
R&S ZNA.
As a prerequisite, the R&S ZNA must accept incoming connections on UDP port 2473.
If necessary, add a corresponding inbound rule to your firewall settings.
Remote command:
n/a
Add Device
Opens the "Add Generic Device" dialog that allows you to configure the connection to
a generic device that was not auto-detected. The configured device is added to the list
of "Configured Devices".
The connection is defined with the following settings:
"Interface" selects an interface/protocol type for the connection. In addition to the
VXI-11, SOCKET, and GPIB interfaces, the analyzer supports any "Other" interface
supported by the installed VISA library.
USB-VISA devices are auto-detected and hence this interface type does not show
up here.
"Address" contains the address for the current interface type. GPIB addresses
must be unique for all devices connected to the GPIB bus (range: 0 to 30), GPIB
and IP addresses must agree with the entries in the VISA library. The remaining
interface types require composite address formats; see Table 5-13.
"Identify" sends an identification query ("IDN?") to the specified device address to
identify the model and serial number of the connected device.